      Axial length acquisition success rates and agreement of four optical biometers and one ultrasound biometer in eyes with dense cataracts

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          To evaluate the axial length acquisition success rates and agreement between various biometric parameters obtained with different biometers in dense cataracts.

          Methods

          Fifty-one eyes were measured using Anterion®, Argos® and IOLMaster® 700 swept-source optical coherence tomography (SS-OCT) biometers, a Pentacam® AXL partial coherence interferometry (PCI) biometer, and an OcuScan® RxP ultrasound biometer. We measured keratometry (K1, flattest keratometry and K2, steepest keratometry), white-to-white (WTW), anterior chamber depth (ACD), lens thickness (LT) and axial length. Cataracts were classified according to the Lens Opacities Classification System III grading system, the dysfunctional lens index (DLI) and Pentacam® nucleus staging (PNS) metrics. Percentage of acquisition success rate and a Bland–Altman analysis for the agreement between biometers were calculated.

          Results

          The mean LOCS III score was 3.63 ± 0.92, the mean DLI was 2.95 ± 1.30 and the mean PNS was 2.36 ± 1.20. The acquisition success rates for the Anterion®, Argos®, IOLMaster® 700, Pentacam® AXL and OcuScan® RxP biometers were 94.12%, 100%, 98.04%, 60.78% and 100%, respectively. There were significant differences in the success rates between biometers ( P = 0.014). There were statistically significant differences between biometers for all parameters evaluated ( P < 0.05). The range of the limit of agreement (LoA) for all comparisons of K1 and K2 were > 1.00 D. The LoA for WTW ranged from 0.095 to 1.050 mm. The LoA for ACD and LT ranged from 0.307 to 0.114 mm and from 0.378 to 0.108 mm, respectively. The LoA for axial length ranged from 0.129 to 2.378 mm.

          Conclusions

          Among optical biometers, those based on SS-OCT technology are more successful at measuring axial length in eyes with dense cataracts.

          Trial Registration: The study was registered with the National Institutes of Health (clinical trial identifier NCT05239715, http://www.clinicaltrials.gov).

          Supplementary Information

          The online version contains supplementary material available at 10.1186/s40662-023-00352-3.

          Related collections

          Most cited references41

          • Record: found
          • Abstract: found
          • Article: not found

          The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group.

          To develop the Lens Opacities Classification System III (LOCS III) to overcome the limitations inherent in lens classification using LOCS II. These limitations include unequal intervals between standards, only one standard for color grading, use of integer grading, and wide 95% tolerance limits. The LOCS III contains an expanded set of standards that were selected from the Longitudinal Study of Cataract slide library at the Center for Clinical Cataract Research, Boston, Mass. It consists of six slit-lamp images for grading nuclear color (NC) and nuclear opalescence (NO), five retroillumination images for grading cortical cataract (C), and five retroillumination images for grading posterior subcapsular (P) cataract. Cataract severity is graded on a decimal scale, and the standards have regularly spaced intervals on a decimal scale. The 95% tolerance limits are reduced from 2.0 for each class with LOCS II to 0.7 for nuclear opalescence, 0.7 for nuclear color, 0.5 for cortical cataract, and 1.0 for posterior subcapsular cataract with the LOCS III, with excellent interobserver agreement. The LOCS III is an improved LOCS system for grading slit-lamp and retroillumination images of age-related cataract.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Calculation of intraocular lens power: a review

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              C constant: new concept for ray tracing-assisted intraocular lens power calculation.

              To evaluate the accuracy of the C constant for ray tracing-assisted intraocular lens (IOL) power calculation.
                Bookmark

                Author and article information

                Contributors
                rdi@oftalvist.es
                Journal
                Eye Vis (Lond)
                Eye Vis (Lond)
                Eye and Vision
                BioMed Central (London )
                2326-0254
                1 September 2023
                1 September 2023
                2023
                : 10
                : 35
                Affiliations
                [1 ]Oftalvist, C/Angel Lozano 11, 03001 Alicante, Spain
                [2 ]GRID grid.5338.d, ISNI 0000 0001 2173 938X, Optics and Optometry and Vision Sciences Department, , University of Valencia, ; Valencia, Spain
                Article
                352
                10.1186/s40662-023-00352-3
                10472586
                37653460
                0474a8ab-2000-4e43-94c1-441ecb127c6c
                © The Author(s) 2023

                Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 9 February 2023
                : 13 July 2023
                Funding
                Funded by: investigator-initiated study grant from Alcon Laboratories
                Award ID: (IIT# 68386009).
                Award Recipient :
                Categories
                Research
                Custom metadata
                © Wenzhou Medical University 2023

                dense cataracts,optical biometry,swept-source optical coherence tomography,partial coherence interferometry,ultrasound biometry

                Comments

                Comment on this article