      Argus II retinal prosthesis implantation with scleral flap and autogenous temporalis fascia as alternative patch graft material: a 4-year follow-up

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          The Argus II retinal prosthesis is composed of an epiretinal electrode array positioned over the macula and connected to an extrascleral electronics case via a silicone cable, running through a sclerotomy. During implantation, the manufacturer recommends to cover the sclerotomy site with a patch of processed human pericardium to prevent postoperative hypotony and conjunctival erosion by the underlying electronics case. Due to biomedical regulations prohibiting the use of this material in France, we developed an alternative technique combining a scleral flap protecting the sclerotomy and an autogenous graft of superior temporalis fascia overlying the electronics case.

          Methods

          The purpose of this study is to describe the 4-year outcomes of this modified procedure in three subjects who underwent Argus II Retinal Prosthesis System implantation. Clinical data consisting of intraocular pressure measurements and tolerance in terms of conjunctival erosion or inflammation were retrospectively assessed over a 4-year postoperative follow-up.

          Results

          None of the three patients implanted with the modified technique developed ocular hypotony over 4 years. A normal, transient conjunctival inflammation occurred during the first postoperative month but conjunctival erosion was not observed in any of the three patients over 4 years. Four years after implantation, the autogenous temporalis fascia graft remained well tolerated and the retinal prosthesis was functional in all three patients.

          Conclusion

          The combination of an autograft of superficial temporalis fascia and a scleral flap efficiently prevented leakage through the sclerotomy site, ocular hypotony, and conjunctival erosion by the extrascleral electronics case. This modified technique is suitable for the implantation of existing and forthcoming retinal prostheses. Superficial temporalis fascia may also be used as alternative to commercial tectonic tissues for scleral wound repair in clinical settings where they are not available.

          Related collections

          Most cited references34

          • Record: found
          • Abstract: found
          • Article: not found

          Interim results from the international trial of Second Sight's visual prosthesis.

          This study evaluated the Argus II Retinal Prosthesis System (Second Sight Medical Products, Inc., Sylmar, CA) in blind subjects with severe outer retinal degeneration. Single-arm, prospective, multicenter clinical trial. Thirty subjects were enrolled in the United States and Europe between June 6, 2007, and August 11, 2009. All subjects were followed up for a minimum of 6 months and up to 2.7 years. The electronic stimulator and antenna of the implant were sutured onto the sclera using an encircling silicone band. Next, a pars plana vitrectomy was performed, and the electrode array and cable were introduced into the eye via a pars plana sclerotomy. The microelectrode array then was tacked to the epiretinal surface. The primary safety end points for the trial were the number, severity, and relation of adverse events. Principal performance end points were assessments of visual function as well as performance on orientation and mobility tasks. Subjects performed statistically better with the system on versus off in the following tasks: object localization (96% of subjects), motion discrimination (57%), and discrimination of oriented gratings (23%). The best recorded visual acuity to date is 20/1260. Subjects' mean performance on orientation and mobility tasks was significantly better when the system was on versus off. Seventy percent of the patients did not have any serious adverse events (SAEs). The most common SAE reported was either conjunctival erosion or dehiscence over the extraocular implant and was treated successfully in all subjects except in one, who required explantation of the device without further complications. The long-term safety results of Second Sight's retinal prosthesis system are acceptable, and most subjects with profound visual loss perform better on visual tasks with system than without it. Copyright © 2012 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found
            Is Open Access

            Long-Term Results from an Epiretinal Prosthesis to Restore Sight to the Blind.

            Retinitis pigmentosa (RP) is a group of inherited retinal degenerations leading to blindness due to photoreceptor loss. Retinitis pigmentosa is a rare disease, affecting only approximately 100 000 people in the United States. There is no cure and no approved medical therapy to slow or reverse RP. The purpose of this clinical trial was to evaluate the safety, reliability, and benefit of the Argus II Retinal Prosthesis System (Second Sight Medical Products, Inc, Sylmar, CA) in restoring some visual function to subjects completely blind from RP. We report clinical trial results at 1 and 3 years after implantation.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              The Detection of Motion by Blind Subjects With the Epiretinal 60-Electrode (Argus II) Retinal Prosthesis.

              To investigate the ability of 28 blind subjects implanted with a 60-electrode Argus II (Second Sight Medical Products Inc) retinal prosthesis system to detect the direction of a moving object. Blind subjects (bare light perception or worse in both eyes) with retinitis pigmentosa were implanted with the Argus II prosthesis as part of a phase 1/2 feasibility study at multiple clinical sites worldwide. The experiment measured their ability to detect the direction of motion of a high-contrast moving bar on a flatscreen monitor in 3 conditions: with the prosthesis system on and a 1-to-1 mapping of spatial information, with the system off, and with the system on but with randomly scrambled spatial information. Fifteen subjects (54%) were able to perform the task significantly better with their prosthesis system than they were with their residual vision, 2 subjects had significantly better performance with their residual vision, and no difference was found for 11 subjects. Of the 15 better-performing subjects, 11 were available for follow-up testing, and 10 of them had significantly better performance with normal rather than with scrambled spatial information. This work demonstrates that blind subjects implanted with the Argus II retinal prosthesis were able to perform a motion detection task they could not do with their native vision, confirming that electrical stimulation of the retina provides spatial information from synchronized activation of multiple electrodes. clinicaltrials.gov Identifier:NCT00407602
                Bookmark

                Author and article information

                Journal
                Clin Ophthalmol
                Clin Ophthalmol
                Clinical Ophthalmology
                Clinical Ophthalmology (Auckland, N.Z.)
                Dove Medical Press
                1177-5467
                1177-5483
                2016
                18 August 2016
                : 10
                : 1565-1571
                Affiliations
                [1 ]INSERM and DHOS, CHNO des Quinze-Vingts
                [2 ]Sorbonne Universités, UPMC Univ Paris 6, Institut de la Vision
                [3 ]INSERM
                [4 ]CNRS, Paris, France
                [5 ]Institute of Ophthalmology, University College London, London, UK
                [6 ]Fondation Ophtalmologique Adolphe de Rothschild
                [7 ]Académie des Sciences, Institut de France, Paris, France
                Author notes
                Correspondence: Pierre-Olivier Barale, CHNO des Quinze-Vingts, 28 rue de Charenton, F-75012 Paris, France, Tel +33 1 40 02 14 04, Fax +33 1 40 02 14 45, Email pobarale@ 123456wanadoo.fr
                Article
                opth-10-1565
                10.2147/OPTH.S112403
                4993412
                93e5b9b6-7c82-49f6-b581-f1acbbde5113
                © 2016 Matet et al. This work is published and licensed by Dove Medical Press Limited

                The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License ( http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.

                History
                Categories
                Original Research

                Ophthalmology & Optometry
                visual prosthesis,retinitis pigmentosa,surgical procedure,conjunctiva,intraocular pressure