      Breakage of a Third Generation Gamma Nail: A Case Report and Review of the Literature

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The use of intramedullary nails to treat trochanteric fractures of the femur has increased with the increasing size of the elderly population. The third generation Gamma nail is currently one of the most popular devices for the treatment of trochanteric fractures. Nail breakage is a rare complication, possibly resulting from fatigue fracture of the implant. We present the first reported case of breakage of a third generation Gamma nail that was not used to treat a pathological fracture. An 83-year-old woman with an unstable trochanteric fracture of the femur was treated using a third generation Gamma nail. She was referred to our hospital 14 months postoperatively with nail breakage at the opening for the lag screw. The breakage was secondary to nonunion, which was thought to be mainly due to insufficient reduction of the fracture. The broken nail was removed, and the patient underwent cemented bipolar hemiarthroplasty. At followup 18 months later, she was mobile with a walker and asymptomatic with no complications. This case shows that inadequate operation such as insufficient reduction of the trochanteric fracture may result in nonunion and implant breakage, even when using a high-strength, well-designed implant.

          Related collections

          Most cited references40

          • Record: found
          • Abstract: found
          • Article: not found

          Nail or plate fixation of intertrochanteric hip fractures: changing pattern of practice. A review of the American Board of Orthopaedic Surgery Database.

          A new method of fixation for intertrochanteric hip fractures that involves the use of an intramedullary nail that interlocks proximally into the femoral head was introduced in the early 1990s. Anecdotal observation of practice patterns during the Part II (oral) American Board of Orthopaedic Surgery examination suggested that the use of this method had increased substantially in recent years in comparison with the more traditional sliding compression screw technique. A study of the Part II database was undertaken to detect changing patterns of care for intertrochanteric fractures. During the process of Board certification, candidates for the Part II (oral) examination submit a six-month surgical case list and patient data into a secure database. The database was searched for all intertrochanteric fractures (International Classification of Diseases, Ninth Revision, code 820.20 or 820.21) over a seven-year period (1999 through 2006). The cases were categorized by intramedullary nail or plate fixation on the basis of surgeon-reported Current Procedural Terminology codes. Relative utilization of the two devices was analyzed according to year and region, and the devices were compared in terms of complications and outcomes. A dramatic change in practice was demonstrated, with the intramedullary nail fixation rate increasing from 3% in 1999 to 67% in 2006. Regional variation was substantial. The highest rate of utilization of intramedullary nail fixation was recorded by candidates from the South, Southeast, and Southwest, who converted to the new technology faster than those in the Northeast, Northwest, and Midwest. Overall, patients managed with plate fixation had slightly less pain and deformity in comparison with those managed with intramedullary nailing, with no significant differences being identified in terms of function or satisfaction. Patients managed with intramedullary nailing had more procedure-related complications, particularly bone fracture. From 1999 to 2006, a dramatic change in surgeon preference for the fixation device used for the treatment of intertrochanteric fractures has occurred among young orthopaedic surgeons. This change has occurred despite a lack of evidence in the literature supporting the change and in the face of the potential for more complications.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures in adults.

            Two types of implants used for the surgical fixation of extracapsular hip fractures are cephalocondylic intramedullary nails, which are inserted into the femoral canal proximally to distally across the fracture, and extramedullary implants (e.g. the sliding hip screw). To compare cephalocondylic intramedullary nails with extramedullary implants for extracapsular hip fractures in adults. We searched the Cochrane Bone, Joint and Muscle Trauma Group Specialised Register (April 2010), The Cochrane Central Register of Controlled Trials (The Cochrane Library 2010, Issue 1), MEDLINE (1950 to March 2010), EMBASE (1980 to 2010 Week 13), and other sources. All randomised and quasi-randomised controlled trials comparing cephalocondylic nails with extramedullary implants for extracapsular hip fractures. Both authors independently assessed trial quality and extracted data. Wherever appropriate, results were pooled. We included 43 trials containing predominantly older people with mainly trochanteric fractures. Twenty-two trials (3749 participants) compared the Gamma nail with the sliding hip screw (SHS). The Gamma nail was associated with increased risk of operative and later fracture of the femur and increased reoperation rate. There were no major differences between implants in wound infection, mortality or medical complications.Five trials (623 participants) compared the intramedullary hip screw (IMHS) with the SHS. Fracture fixation complications were more common in the IMHS group. Results for post-operative complications, mortality and functional outcomes were similar in both groups.Three trials (394 participants) showed no difference in fracture fixation complications, reoperation, wound infection and length of hospital stay for proximal femoral nail (PFN) versus the SHS.None of the 10 trials (1491 participants) of other nail versus extramedullary implant comparisons for trochanteric fractures provided sufficient evidence to establish definite differences between the implants under test.Two trials (65 participants) found intramedullary nails were associated with fewer fracture fixation complications than fixed nail plates for unstable fractures at the level of the lesser trochanter.Two trials (124 participants) found a tendency to less fracture healing complications with the intramedullary nails compared with fixed nail plates for subtrochanteric fractures. With its lower complication rate in comparison with intramedullary nails, and absence of functional outcome data to the contrary, the SHS appears superior for trochanteric fractures. Further studies are required to confirm whether more recently developed designs of intramedullary nail avoid the complications of previous nails. Intramedullary nails may have advantages over fixed angle plates for subtrochanteric and some unstable trochanteric fractures, but further studies are required.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Gamma nails revisited: gamma nails versus compression hip screws in the management of intertrochanteric fractures of the hip: a meta-analysis.

              Concerns about the Gamma nail have largely been fueled by early randomized trials and meta-analyses suggesting an increased risk of subsequent femoral shaft fractures when compared with compression hip screws. Whereas meta-analyses favor compression hip screws over first-generation Gamma nails, little is known as to whether the newer Gamma nail designs and the improved learning curve associated with the implants have reduced the risk of femoral shaft fracture. The current meta-analysis aimed to explore the effects of time and Gamma Nail design on the risks of femoral shaft fracture after treatment of extracapsular hip fractures. We searched computerized databases (Medline, Cochrane, and SciSearch) for published randomized clinical trials from 1969 to 2002, and we identified additional studies through hand searches of major orthopedic journals, bibliographies of major orthopedic texts, and personal files. Two investigators independently graded study quality and abstracted relevant data. We abstracted information on subsequent femoral shaft fracture rates from studies. We pooled data using a random-effects model and tested for heterogeneity using the I test. We conducted sensitivity analyses by date and by generation of the Gamma nail. We further conducted a cumulative meta-analysis to explore the treatment effect over time. We identified 25 relevant randomized trials from 1991 to 2005. In earlier studies (N = 1585 patients), Gamma nails increased the risk of femoral shaft fracture 4.5 times compared with a compression hip screw (95% confidence interval: 1.78-11.36, P = 0.0014, I(2) = 0%). However, among the most recent studies (2000-2005), Gamma nails did not significantly increase femoral shaft fracture risk (relative risk = 1.65, 95% confidence interval: 0.50-5.44, P = 0.41, I(2) = 0%). The most recent study (2005) found no difference in femoral fracture rates (relative risk = 1.03, 95% confidence interval = 0.06-16.2, P = 0.99). Our meta-analysis of randomized trials suggests that previous concerns about increased femoral shaft fracture risk with Gamma nails have been resolved with improved implant design and improved learning curves with the device. Earlier meta-analyses and randomized trials should be interpreted with caution in light of more recent evidence.
                Bookmark

                Author and article information

                Journal
                Case Rep Orthop
                Case Rep Orthop
                CRIM.ORTHOPEDICS
                Case Reports in Orthopedics
                Hindawi Publishing Corporation
                2090-6749
                2090-6757
                2013
                20 May 2013
                : 2013
                : 172352
                Affiliations
                Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe 650-0017, Japan
                Author notes

                Academic Editors: M. Cadossi, L. G. Grossterlinden, and D. S. Karataglis

                Author information
                https://orcid.org/0000-0003-0951-5812
                https://orcid.org/0000-0002-0435-8398
                Article
                10.1155/2013/172352
                3671522
                23762698
                53c4dfb8-9023-4c7a-9fd5-0f83c453a3c4
                Copyright © 2013 Takashi Iwakura et al.

                This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 11 March 2013
                : 8 May 2013
                Categories
                Case Report

                Orthopedics
                Orthopedics

                Comments

                Comment on this article