0
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: not found

      Optical Coherence Tomography-Based Modeling of Stent Deployment in Heavily Calcified Coronary Lesion.

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          In this work, a heavily calcified coronary artery model was reconstructed from optical coherence tomography (OCT) images to investigate the impact of calcification characteristics on stenting outcomes. The calcification was quantified at various cross sections in terms of angle, maximum thickness, and area. The stent deployment procedure, including the crimping, expansion, and recoil, was implemented. The influence of calcification characteristics on stent expansion, malapposition, and lesion mechanics was characterized. Results have shown that the minimal lumen area following stenting occurred at the cross section with the greatest calcification angle. The calcification angle constricted the stretchability of the lesion and thus resulted in a small lumen area. The maximum principal strain and von Mises stress distribution patterns in both the fibrotic tissue and artery were consistent with the calcification profiles. The radially projected region of the calcification tends to have less strain and stress. The peak strain and stress of the fibrotic tissue occurred near the interface with the calcification. It is also the region with a high risk of tissue dissection and strut malapposition. In addition, the superficial calcification with a large angle aggregated the malapposition at the middle of the calcification arc. These detailed mechanistic quantifications could be used to provide a fundamental understanding of the role of calcification in stent expansions, as well as to exploit their potential for enhanced pre- and post-stenting strategies.

          Related collections

          Author and article information

          Journal
          J Biomech Eng
          Journal of biomechanical engineering
          ASME International
          1528-8951
          0148-0731
          May 01 2020
          : 142
          : 5
          Affiliations
          [1 ] Department of Mechanical and Materials Engineering, University of Nebraska-Lincoln, Lincoln, NE 68588.
          [2 ] Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH 44106.
          [3 ] Department of Medicine-Cardiovascular Medicine, Cardiovascular Imaging Core Laboratory, Harrington Heart & Vascular Institute, University Hospitals Cleveland Medical Center, Cleveland, OH 44106.
          [4 ] Department of Mechanical and Materials Engineering, University of Nebraska-Lincoln, Lincoln, NE 68588; Department of Biomedical and Chemical Engineering and Sciences, Florida Institute of Technology, Melbourne, FL 32901.
          Article
          1066040
          10.1115/1.4045285
          7104774
          31654052
          0d96c51b-1ed1-4e0e-b6b8-6e2006156c9f
          History

          finite element analysis,malapposition,stent deployment,optical coherence tomography (OCT),stent underexpansion,calcification,percutaneous coronary intervention

          Comments

          Comment on this article